Development of preanalytical criteria assessment of quality of vacuum blood collection systems affecting clinical examination of patients

https://doi.org/10.15789/1563-0625-DOP-1920

Abstract

This paper presents comparative studies of preanalytical quality indices of vacuum gel systems (VGS) from four manufacturers, i.e., GL 795 (China), Vacuette (Austria), BD Vacutainer SST (USA) and Zdravmedtech (Russia). The analysis of the number of qualitative defects of the preanalytic phase of blood collection in vacuum gel tubes at the inpatient departments and in Department of Clinical Chemistry (preanalytic laboratory), in accordance with adopted criteria based on Russian industry standards, which are implemented with European indices of pre-analytic stage. The obtained results allowed us to identify a number of qualitative defects at pre-analytic stage for the vacuum gel systems from each manufacturer, including to summarize the data on number of these defects. Based on these results, the most common defects in each of the VGS presented by manufacturers are formulated. The results obtained can be used by manufacturers to improve the quality (structures) of the mentioned vacuum gel tubes.

The results of the comparative characteristics of VGS of different manufacturers were evaluated by their influence on the indexes of hemolysis (IH), ictericity, lipemia. The resulting indices of ictericity and lipemia do not allow to use them as criteria for assessing quality of the gel systems. The frequency of IH occurrence in serum at > 50 conventional units in VGS of different manufacturers was studied. It was revealed, that IH in the range from 10 to 50 conventional units and more, can be used as a criterion of VGS quality.

Comparison of routine biochemical parameters in VGS of different manufacturers, i.e., alanine aminotransferase, aspartate aminotransferase, direct bilirubin, lactate dehydrogenase (LDH), serum K+, was also carried out. It was found that the largest deviations from the average value were obtained for aspartate aminotransferase indices of more than 12.47% in Russian test tubes, and for direct bilirubin of more than 12.25% in Chinese test tubes and more than 9.15% in Russian test tubes. The study showed influence of IH upon quantitative values of lactate dehydrogenase for VGS from different manufacturers. It is proposed to use lactate dehydrogenase as an analyte. On this basis, it is possible to recommend a comparative quality assessment of vacuum gel systems, with IH cutoff level of more than 19 conventional units. It remains relevant to search for other analytes that can be used to assess the quality of VGS from different manufacturers. Some areas of research in biochemical testing are outlined. Preliminary data are obtained on usage of vacuum systems from different manufacturers for quality assessment, as well as conditions for performing these tests, along with the level of the hemolysis index of more than 10 conventional units.

About the Authors

S. P. Kazakov
N. Burdenko Main Military Clinical Hospital; Russian Medical Academy of Continuing Professional Education
Russian Federation

Kazakov Sergey P. – PhD, MD (Medicine), Associate Professor, Head, Сenter for Clinical Laboratory Diagnostics, N. Burdenko Main Military Clinical Hospital; Head, Department of Medical Biochemistry and Immunopathology, Russian Medical Academy of Continuing Professional Education

105094, Moscow, Gospitalnaya Square, 3
Phone: 7 (499) 263-01-73



L. A. Suslova
N. Burdenko Main Military Clinical Hospital
Russian Federation

Head, Quality Control Laboratory, Сenter for Clinical Laboratory Diagnostics

Moscow



References

1. Bugrov A.V., Dolgov V.V., Kazakov S.P., Lugovskaya S.A., Mironova I.I., Pochtar M.E., Rakova N.G., Roitman A.P., Romanova L.A., Selivanova A.V., Sosnin D.Yu., Shabalova I.P., Shevchenko O.P., Schetnikovich K.A. Clinical laboratory diagnostics: a textbook. Moscow: Labdiag, 2017. 464 p.

2. Clinical laboratory technologies. Quality assurance of clinical laboratory tests. Part 4. Rules for conducting of preanalytical stage.

3. State standard ISO-6710-2011 Single-use containers for venous blood specimen collection. Technical requirements and test methods (state standard is put into effect by the order of Federal Agency for technical regulation and metrology of December 13, 2011 N1379-st.

4. State standard ISO 10993-4-2011 Medical devices. Biological evaluation of medical devices. Part 4. Selection of tests for interactions with blood.

5. Guder V.G., Naraianan S., Visser G., Cavta B. Diagnostic samples: from the patient to the lab. Moscow: Labora, 2010. 117 p.

6. Dymova O.V. Quality indicators at different stages of laboratory research]. Spravochnik zaveduyuschego CLD = Handbook of the Head of the Clinical Diagnostic Laboratory, 2018, no. 9, pp. 16-24. (In Russ.)

7. Klimenkova O.A., Berestovskaya V.S., Laricheva E.S. The index of hemolysis: from discussion to decision by the breakdown of preanalytical quality]. Sovremennaya laboratoriya = Modern Laboratory, 2013, Vol. 3, no. 16, pp. 38-40. (In Russ.)

8. Klimenkova O.A., Emanuel A.V. Quality indicators: the cjnditions for the benchmarking of laboratory services]. Vestnik roszdravnadzora = Bulletin of Roszdravnadzor, 2014, no. 1, pp. 14-22. (In Russ.)

9. Klimenkova O.A., Berestovskaya V.S., Ivanov G.A., Emanuel A.V. Experience with the use of the hemolysis index as indicator of preanalytical quality for immunochemical studies]. Vestnik roszdravnadzora = Bulletin of Roszdravnadzor, 2014, no. 3, pp. 33-38. (In Russ.)

10. Klimenkova O.A. The management of quality in preanalytical phase using hemolysis index: abstract dis. ... cand. med. sci. St. Petersburg, 2019. 25 p.

11. Kudryashov S.R., Kanischev Yu.N., Putkov S.B., Esaulenko N.B., Karpov V.O., Ovcharenko V.P., Izgorodin A.S., Jukova E.E., Suslova L.A., Parshakova E.V. Instructions for the preanaiytic stage (the procedure for taking, storing and transporting) with biomaterial for laboratory research in the center of clinical laboratory diagnostics of the MMCH named after academician N.N. Burdenko: methodological guidance. Moscow: Eco-Press, 2016. 216 p.

12. Menshikov V.V. Clinical patient safety and reliability of laboratory information (lecture)]. Clinicheskaya laboratornaya diagnostika = Clinical Laboratory Diagnostics, 2013, no. 6, pp. 29-36. (In Russ.)

13. Menshikov V.V., Emanuel A.V., Pervushin Yu.V., Cvirenko S.V. Ensuring clinical safety of obtaining and using laboratory information: clinical recommendations. Moscow: Association of specialists and organizations of laboratory service “Federation of laboratory medicine”, 2013.

14. Moshkin A.V. Assessment of the degree of hemolysis on the biochemical analyzer VITROS 5.1FS – a possible indicator of the quality of sampling and transportation]. Laboratoriya = Laboratory, 2011, no. 3, pp. 18-19. (In Russ.)

15. Khorovskaya L.A., Shmidt I.O., Kovalevskaya S.N. Comparison of results of biochemical blood tests of patients using vacuum tubes Lind-vac (Estonia) and Greiner (Austria)]. Spravochnik zaveduyuschego CLD = Handbook of the Head of the Clinical Diagnostic Laboratory, 2015, no. 2, pp. 9-19. (In Russ.)

16. Khorovskaya L.A., Danilova D.G., Shmidt I.O. Experience in Validation of UNIVAC vacuum tubes in a medical laboratory]. Spravochnik zaveduyuschego CLD = Handbook of the Head of the Clinical Diagnostic Laboratory, 2017, no. 9, pp. 3-16. (In Russ.)

17. Gawria G., Tillmar L., Landberg E. A comparison of stability of chemical analytes in plasma from the BD TM Vacutainer® Barricor tube with mechanical separator versus tubes containing gel separator. J. Clin. Lab. Anal., 2020, Vol. 34, no. 2, e23060. doi: 10.1002/jcla.23060.

18. Kavsak P.A., Mansour M., Wang L., Campeau S., Clark L., Brooks D., Trus M. Assessing pneumatic tube systems with patient-specific population and laboratory derived criteria. Clin. Chem., 2012, Vol. 58, pp. 792-795.

19. Plebani M., Chiozza M.L., Sciacovelli L. Towards harmonization of quality indicators in laboratory medicine. Clin. Chem. Lab. Med., 2013, Vol. 51, no. 1, рp. 187-195.

20. Simundic A.M., Bölenius K., Cadamuro J., Church S., Cornes M.P., van Dongen-Lases E.C., Eker P., Erdeljanovic T., Grankvist K., Guimaraes J.T., Hoke R., Ibarz M., Ivanov H., Kovalevskaya S., Kristensen G.B.B., Lima-Oliveira G., Lippi G., von Meyer A., Nybo M., de la Salle B., Seipelt C., Sumarac Z., Vermeersch P. Joint EFLM-COLABIOCLI Recommendation for venous blood sampling. Clin. Chem. Lab. Med., 2018, Vol. 56, no. 12, pp. 2015-2038.

21. Sztefko K. Immunodiagnostics and patient safety. De Gruyter, 2011. 162 p.


Supplementary files

1. Metadanie
Subject
Type Исследовательские инструменты
Download (15KB)    
Indexing metadata ▾
2. Подписи авторов
Subject
Type Исследовательские инструменты
Download (462KB)    
Indexing metadata ▾
3. Титульный лист
Subject
Type Исследовательские инструменты
Download (17KB)    
Indexing metadata ▾
4. Резюме
Subject
Type Исследовательские инструменты
Download (16KB)    
Indexing metadata ▾
5. Название рисунков
Subject
Type Исследовательские инструменты
Download (10KB)    
Indexing metadata ▾
6. Рисунок 1
Subject
Type Исследовательские инструменты
Download (114KB)    
Indexing metadata ▾
7. Рисунок 2
Subject
Type Исследовательские инструменты
Download (26KB)    
Indexing metadata ▾
8. Список литературы
Subject
Type Исследовательские инструменты
Download (32KB)    
Indexing metadata ▾
9. Таблица 1
Subject
Type Исследовательские инструменты
Download (14KB)    
Indexing metadata ▾
10. Таблица 2
Subject
Type Исследовательские инструменты
Download (16KB)    
Indexing metadata ▾
11. Таблица 3
Subject
Type Исследовательские инструменты
Download (18KB)    
Indexing metadata ▾
12. Таблица 4
Subject
Type Исследовательские инструменты
Download (12KB)    
Indexing metadata ▾
13. Таблица 5
Subject
Type Исследовательские инструменты
Download (14KB)    
Indexing metadata ▾
14. Таблица 6
Subject
Type Исследовательские инструменты
Download (11KB)    
Indexing metadata ▾
15. Конфликт интересов
Subject
Type Исследовательские инструменты
Download (224KB)    
Indexing metadata ▾
16. Резюме2
Subject
Type Other
Download (15KB)    
Indexing metadata ▾
17. Resume2
Subject
Type Other
Download (15KB)    
Indexing metadata ▾
18. РАЗРАБОТКА ПРЕАНАЛИТИЧЕСКИХ КРИТЕРИЕВ ОЦЕНКИ КАЧЕСТВА ВАКУУМНЫХ СИСТЕМ СБОРА КРОВИ, ВЛИЯЮЩИХ НА КЛИНИЧЕСКОЕ ОБСЛЕДОВАНИЕ ПАЦИЕНТОВ
Subject Дополнительные файлы к статье после рецензии
Type Резюме(исправленное), статья(исправленное) таблица 1 (исправленное)
Download (16KB)    
Indexing metadata ▾
19. РАЗРАБОТКА ПРЕАНАЛИТИЧЕСКИХ КРИТЕРИЕВ ОЦЕНКИ КАЧЕСТВА ВАКУУМНЫХ СИСТЕМ СБОРА КРОВИ, ВЛИЯЮЩИХ НА КЛИНИЧЕСКОЕ ОБСЛЕДОВАНИЕ ПАЦИЕНТОВ
Subject Дополнительные файлы к статье после рецензии
Type Резюме(исправленное), статья(исправленное) таблица 1 (исправленное)
Download (16KB)    
Indexing metadata ▾
20. РАЗРАБОТКА ПРЕАНАЛИТИЧЕСКИХ КРИТЕРИЕВ ОЦЕНКИ КАЧЕСТВА ВАКУУМНЫХ СИСТЕМ СБОРА КРОВИ, ВЛИЯЮЩИХ НА КЛИНИЧЕСКОЕ ОБСЛЕДОВАНИЕ ПАЦИЕНТОВ
Subject Дополнительные файлы к статье после рецензии
Type Резюме(исправленное), статья(исправленное) таблица 1 (исправленное)
Download (16KB)    
Indexing metadata ▾
21. РАЗРАБОТКА ПРЕАНАЛИТИЧЕСКИХ КРИТЕРИЕВ ОЦЕНКИ КАЧЕСТВА ВАКУУМНЫХ СИСТЕМ СБОРА КРОВИ, ВЛИЯЮЩИХ НА КЛИНИЧЕСКОЕ ОБСЛЕДОВАНИЕ ПАЦИЕНТОВ
Subject Дополнительные файлы к статье после рецензии
Type Резюме(исправленное), статья(исправленное) таблица 1 (исправленное)
Download (17KB)    
Indexing metadata ▾
22. РАЗРАБОТКА ПРЕАНАЛИТИЧЕСКИХ КРИТЕРИЕВ ОЦЕНКИ КАЧЕСТВА ВАКУУМНЫХ СИСТЕМ СБОРА КРОВИ, ВЛИЯЮЩИХ НА КЛИНИЧЕСКОЕ ОБСЛЕДОВАНИЕ ПАЦИЕНТОВ
Subject Дополнительные файлы к статье после рецензии
Type Резюме(исправленное), статья(исправленное) таблица 1 (исправленное)
Download (438KB)    
Indexing metadata ▾

Review

For citations:


Kazakov S.P., Suslova L.A. Development of preanalytical criteria assessment of quality of vacuum blood collection systems affecting clinical examination of patients. Medical Immunology (Russia). 2020;22(3):585-598. (In Russ.) https://doi.org/10.15789/1563-0625-DOP-1920

Views: 1316


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1563-0625 (Print)
ISSN 2313-741X (Online)